首页 | 官方网站   微博 | 高级检索  
     


Negative‐pressure‐induced collector for a self‐balance free‐flow electrophoresis device
Authors:Qiang Zhang  Chen‐Gang Guo  Fan‐Zhi Kong  Cheng‐Xi Cao  Liu‐Yin Fan  Xin‐Qiao Jin
Affiliation:1. Laboratory of Analytical Biochemistry and Bioseparation, Key State Laboratory of Microbial Metabolism, School of Life Science and Biotechnology, Shanghai Jiao Tong University, Shanghai, China;2. School of Mechanical Engineering, Institute of Refrigeration and Cryogenics, Shanghai Jiao Tong University, Shanghai, China
Abstract:Uneven flow in free‐flow electrophoresis (FFE) with a gravity‐induced fraction collector caused by air bubbles in outlets and/or imbalance of the surface tension of collecting tubes would result in a poor separation. To solve these issues, this work describes a novel collector for FFE. The collector is composed of a self‐balance unit, multisoft pipe flow controller, fraction collector, and vacuum pump. A negative pressure induced continuous air flow rapidly flowed through the self‐balance unit, taking the background electrolyte and samples into the fraction collector. The developed collector has the following advantages: (i) supplying a stable and harmonious hydrodynamic environment in the separation chamber for FFE separation, (ii) effectively preventing background electrolyte and sample flow‐back at the outlet of the chamber and improving the resolution, (iii) increasing the preparative scale of the separation, and (iv) simplifying the operation. In addition, the cost of the FFE device was reduced without using a multichannel peristaltic pump for sample collection. Finally, comparative FFE experiments on dyes, proteins, and cells were carried out. It is evident that the new developed collector could overcome the problems inherent in the previous gravity‐induced self‐balance collector.
Keywords:Flow field  Free‐flow electrophoresis  Sample collection
设为首页 | 免责声明 | 关于勤云 | 加入收藏

Copyright©北京勤云科技发展有限公司    京ICP备09084417号-23

京公网安备 11010802026262号